After-sales Service: | Following Updating |
---|---|
Warranty: | Under Discussion |
Weighing: | 5000kg |
Still deciding? Get samples of $ !
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Operating conditions |
The ambient temperature is -20 ºC~60 ºC, avoiding direct sunlight, and the temperature change rate is less than 10 ºC/h |
Environmental humidity of 20% to 80% RH (temperature of 5-40 ºC), |
10%~50% RH (temperature 40-50 ºC), no condensation |
The operating environment is free of flammable, explosive, toxic, water mist, or corrosive media |
Vibration frequency range: 10-55Hz |
Displacement amplitude: 0.035mm |
Impact not allowed |
Magnetic field ≤ 400A/m (DC and 50Hz) |
Common mode interference common mode voltage of 250VACrms or lower, 50Hz/60Hz |
The common mode voltage between input channels is 250VACrms or lower, 50Hz/60Hz |
Serial mode interference power range or 50mVACrms, 50Hz/60Hz |
Preheating time: At least 30 minutes after turning on the power |
Installation location indoor |
The maximum installation angle can be tilted backwards by 30 °, and left and right are horizontal |
Altitude < 2000m |
source |
Rated power supply voltage 180VAC~240VAC |
Allow power supply voltage range of ± 10% |
Rated power frequency 50Hz/60Hz |
Power frequency allowable range ± 2% |
Maximum power consumption 30VA (including optional features) |
Power frequency with an allowed interruption time of less than 1 week |