Bridge resistance (full bridge): foil strain gauge 350 Ω, nominal
Excitation:
5 V DC
Nominal sensitivity:
1 mV/V, nominal*
Insulation resistance:
> 10 MΩ
Calibration resistor:
100 kΩ ± 0.1 %
The bridge output voltage, resulting from a shunt of this value,
is shown in the calibration certificate.
*Deviations from the stated value are possible.
Operating temperature:
0 °C ... + 80 °C
Nominal temperature range:
+ 15 °C ... + 70 °C
Influence of temperature on zero:
≤ ± 1.50 % F.S./50 K
Influence of temperature on sensitivity:
≤ + 1.50 % Rdg./50 K
Non-linearity:
measuring range
≤ 0 ... 2000 N
< 0.5 % F.S.
measuring range 0 ... 5000 N
< 0.75 % F.S.
Hysteresis:
measuring range
≤ 0 ... 2000 N
< 0.2 % F.S.
measuring range 0 ... 5000 N
< 0.3 % F.S.
Non-repeatability on unchanged mounting position:
< 0.2 % F.S.
Deflection, full scale:
approx. 30 µm
Static overload safe:
150 % of capacity
Dynamic performance:
recommended
50 % of capacity
maximum
70 % of capacity
Material:
High-grade stainless steel 1.4542
Electrical connection:
shielded, TPE coated cable with bare ends for soldering,
length approx. 2 m, bending radius ≥ 10 mm
Protection class:
acc. to EN 60529
IP54
Wiring code:
white
excitation voltage
positive
brown
excitation voltage
negative
yellow
signal output
positive
green
signal output
negative
Dimensions:
see table and scale drawing
General tolerances of dimensioning:
acc. to ISO 2768-f
Weight: