Bridge resistance:
measuring range ≤ 0 ... 200 N
500 Ω, nominal*
measuring range
≥ 0 ... 500 N
350 Ω, nominal*
Excitation:
5 V DC
Nominal value:
measuring range ≤ 0 ... 200 N
5 ... 30 mV/V, nominal*
measuring range ≥ 0 ... 500 N
1.2 mV/V, nominal*
Insulation resistance:
> 10 MΩ
*Deviations from the stated value are possible.
Nominal temperature range:
measuring range ≤ 0 ... 200 N
+ 15 °C ... + 60 °C
measuring range
≥ 0 ... 500 N
+ 15 °C ... + 70 °C
Range of operating temperature:
0 °C ... + 80 °C
Influence of temperature on zero:
measuring range ≤ 0 ... 200 N
≤ ± 2.5 % F.S./50 K
measuring range ≥ 0 ... 500 N
≤ ± 1.5 % F.S./50 K
Influence of temperature on sensitivity:
measuring range ≤ 0 ... 200 N
≤ ± 2.5 % Rdg./50 K
measuring range ≥ 0 ... 500 N
≤ ± 1.5 % Rdg./50 K
Combined value consisting of non-linearity, hysteresis and non
repeatability, in installation position:
measuring range ≤ 0 ... 500 N
< ± 0.9 % F.S.
measuring range ≥ 0 ... 1000 N
< ± 0.5 % F.S.
Kind of measurement:
compressive and tensile forces
calibration in tensile direction (preferential direction)
Upon operation against the preferential direction a changed
characteristic is possible.
Deflection, full scale:
max. 60 µm
Static overload safe:
100 % of capacity
Overload:
200 % of capacity
Dynamic performance: recommended
50 % of capacity
maximum
70 % of capacity
Material:
1.4542
Electrical connection: Measuring range ≤ 0 ... 50 N shielded,
PTFE coated cable with an open end for soldering. Circuit board
(70 x 8 mm) with balance resistors 30 cm away from the cable's end.
Range ≤ 0 ... 500 N: shielded, TPE coated cable with an open end for
soldering.
Cable length:
2 m
Bending radius:
30 mm
Protection class:
acc. to EN 60529
IP54
Wiring code:
white
excitation voltage
positive
brown
excitation voltage
negative
green
signal output
negative
yellow
signal output
positive
Dimensions:
refer to drawing
General tolerance of dimensioning:
acc. to ISO 2768-f
Weight:
refer to table