Bridge resistance (full bridge):
measuring ranges ≤ 0 ... 5 N semiconductor 500 Ω, nominal
measuring ranges ≥ 0 ...10 N foil
350 Ω, nominal
Excitation:
5 V DC
Nominal value:
refer to table
Insulation resistance:
> 5000 MΩ by 50 V DC
Shunt calibration resistor:
59 kΩ ± 0.1 %
The bridge output voltage caused by a shunt of this value is shown
in the calibration certificate.
Range of operating temperature:
- 55 °C ... + 120 °C
Nominal temperature range:
+ 15 °C ... + 70 °C
Influence of temperature on zero:
≤ ± 0.02 % F.S./K
Influence of temperature on sensitivity: < + 0.02 % Rdg./K
Non-linearity:
measuring ranges
≤ 0 ... 5 N
< ± 0.5 % F.S.
measuring ranges
≥ 0 ...10 N
< ± 0.25 % F.S.
Accuracy:
measuring ranges
≤ 0 ... 5 N
< ± 0.5 % F.S.
measuring ranges
≥ 0 ...10 N
< ± 0.25 % F.S.
Non-repeatability:
< ± 0.1 % F.S.
Deflection full scale:
measuring ranges
≤ 0 ... 5 N 13 µm ... 38 µm
measuring ranges
≥ 0 ...10 N 25 µm ... 76 µm
Static overload capacity:
150 % of nominal load
Maximum static overload stop:
500 % of nominal load
Dynamic load: recommended
70 % of nominal load
maximum
100 % of nominal load
Material:
stainless steel 17-4 PH (similar to 1.4542)
Electrical connection
Highly flexible teflon isolated with open ends for soldering. Length
approx. 1.5 m. Steep board, width approx. 7 mm, length 50 mm, for
bridge balance, calibration and temperature compensation approx.
0.6 m away from the sensor body. Cable shielding between sensor
and circuit board.
Protecting class: acc. to EN 60529
IP54
Wiring code:
red excitation voltage
positive
black excitation voltage
negative
green signal output
negative
white signal output
positive
Dimensions: refer to table and dimensional drawing
Weight:
refer to table