1.Adopting large dot-matrix LCD, it can display the dissolved oxygen and temperature values of the measured water samples at the same time; |
2.Support zero oxygen calibration and fullness calibration, manual barometric pressure compensation and automatic temperature compensation; |
3.With (4~20) mA DC isolated signal output, the upper limit and lower limit of output can be set arbitrarily in the range of (0~200) μg/L; |
4.With RS-232 output interface; |
5.Support automatic upper and lower limit alarm function and power failure protection function. |
Model | SJG-9435B | |
Measuring range | Dissolved oxygen | (0~19.9)µg/L, (20~199)µg/L, and (0.20~20.00) mg/L |
Temperature | (-10.0~120.0)°C | |
Resolution | Dissolved oxygen | 0.1µg/L; 1µg/L; 0.01mg/L |
Temperature | 0.1ºC | |
Instrument basic error | Dissolved Oxygen | ±5%(FS) |
Temperature | ±0.5ºC(0.0ºC≤T≤60.0ºC);±1.0ºC(Others) | |
Basic error of electronic unit | Dissolved oxygen | ±3%(FS) |
Temperature | ±0.3ºC | |
Temperature compensation error of electronic unit | ±0.3%(FS)/ °C | |
Stability of electronic unit | ±2%(FS)/24h | |
90% Response Time | 2min(25ºC) | |
Instrument protection grade | IP65 | |
Power supply | AC (220±22)V, (50±1)Hz; or DC 24V | |
Sensor cable length (m) | 5 | |
Dimensions (mm), Weight (kg) | 166×166×110, 2 | |
Opening size (mm) | 138×138 | |
Inlet pipe diameter/drainage pipe diameter | Outer diameter Φ6 |